The PCU-2000 Pressure
Control unit is a two-channel, patient-isolated,
line-powered amplifier/interface unit for
connecting one or two Millar Mikro-Tip catheters
to a MDD approved pressure monitor and data
acquisition system. Each channel has an
electrically isolated input for patient safety.
The analog output signal level of 1V/100 mmHg is
compatible with most pressure monitors and data
acquisition systems including AD Instruments
PowerLab.
The PCU-2000
provides electrically isolated bridge excitation
and signal amplification for each channel as
well as separate balance, or zero, controls.
Push button switches provide calibration signals
equivalent to 0, 25, 100 and 125 mmHg on each
channel. A Standby/Transducer switch allows the
operator to reset zero and recalibrate during a
procedure with the transducer in place in the
subject.
LED bar graphs visually
confirm transducer functionality. Pressure input
connectors are color-coded Redel style, while
pressure output connectors are 1/4-inch phone
plug style. Pressure output cables are sold
separately.
-
CE approved
-
Patient isolated
-
Supports one
dual-sensor or two single-sensor Mikro-Tip
catheters
-
Line powered

Mikro-Tip Pressure
Catheters are provided with either Viking
(4-pin) or Millar Low Profile (4-pin) connectors
depending on the model purchased. Catheter
models with Viking connectors may connect
directly to the Pressure Control Units.
Connecting the catheter with a Viking connector
directly to the control unit limits access to
the subject or patient. It is recommended that
an interface cable (TEC-10C) be used between the
catheter and control unit. All catheters with
low profile pressure connectors require an
appropriate interface cable (TEC-10D) to the
control unit. Each pressure channel requires a
separate interface cable and control unit.
 |
|

Single Pressure Sensor
Dual Sensor
Angio Catheter
Pressure Volume
Catheters
MPVS PV
Conductance System
Single Pressure
PCU-2000
Urodynamic
Catheters


|