Developed
specifically for the professional, the next
generation MicroLab employs Micro Medical’s
acclaimed precision Gold Standard Digital Volume
Transducer which is especially suited to
measuring very low flow rates in patients with
COPD.
All the unit’s
features can be easily accessed by simply
touching the chosen function’s corresponding
icon. This, together with fast textual data
entry using a stylus or the optional MicroMouse
makes the system’s use highly intuitive.
A quiet and very fast printer with easy-load
paper mechanism is integrated. The printout can
be fully customised and can also be linked
directly to external printers.
Equipped with many
advanced functions the MicroLab is fully
compatible with Spida 5 and SpidaXpert the
worlds’ most comprehensive Spirometry PC
software packages and meets all International
standards for performance and accuracy. There
are over 40,000 Micro Medical MicroLab
Spirometers in use around the world; this new
generation model is truly the most advanced
portable Spirometer.
Key Features:-
-
High definition colour touch-screen
-
Simple screen
icon interface using stylus or optional
MicroMouse
-
Fast on-screen
textual entry enabling comments and report
writing
-
Fast, quiet
printer with easy load paper mechanism
-
Fully
customisable printout format
-
Direct
connection to external printers.( HP Deskjet
6840 HP Business Inkjet 1000, HP Officejet
Pro.
-
Unique
contextual Help Screens (language
selectable) with dedicated Help Button
-
41 inspiratory
and expiratory indices measurable
-
On-screen
graphical child incentive device
-
Large 2000
patient test memory
-
Optional Spida 5
or SpidaXpert PC software
-
Internationally
recognised on-screen test quality assurance
prompts (e.g. slow start, abrupt end etc.)
ensuring quality test results
-
Unique ‘today’s
testing feature’ - creating a list of
patients tested or to be tested
-
Numerous
predicted values and languages selectable
-
Lung Age’
assessment and textual result interpretation
by ATS (9), BTS-NICE (10) or Enright (11)
-
Body Mass Index
and Dyspnoea Score
-
Bronchodilator
and/or Steroid assessment
-
Complete with
all accessories in a sturdy carrying case
Technical Specifications
|
Specification |
|
Measurements
Forced: |
VC, FEV.75,
FEV1, FEV3, FEV6, FVC, PEF, FEV.75/VC,
FEV.75/FVC, FEV1/VC, FEV1/FVC (FER),
FEV3/VC, FEV3/FVC, FEV.75/FEV6,
FEV1/FEV6, FEF25 (MEF75), FEF50 (MEF50),
FEF75 (MEF25), FEF25-75 (MMEF), FEF50/VC,
FEF50/FVC,
MMEF/FVC (FEF25-75/FVC), FIV1, FIVC, PIF,
FIV1/FIVC (FIR), FIF25 (MIF75), FIF50 (MIF50),
FIF75 (MIF25), R50 (FEF50/FIF50), MET25-75,
FET, MVV (ind)
|
|
Measurements
(relaxed): |
EVC, IVC,
IC,VT (TV), ,Ti,Te,Ti/Ttot.,VT/Ti
(TV/Ti), IRV, ERV, FR |
|
Tests per
subject: |
5 relaxed VC
manoeuvres and 8 forced manoeuvres for
each baseline and two post examinations |
|
Predicted
Values: |
Various –
depends upon national preference
|
|
Transducer:
|
Micro
Medical Gold Standard Bi-Directional
Digital Volume |
|
Resolution:
|
10ml volume
0.03l/s flow |
|
Accuracy:
|
+/-3%.To ATS
recommendations – Standardisation of
spirometry 1994 update for flows and
volumes |
|
Storage:
|
2000
patients’ tests including Flow/Volume
loops and Volume/Time curves
|
|
Printer
Output: |
HP Deskjet
6840, HP Business Inkjet 1000
HP Officejet Pro K550 |
|
Printer
Speed (Internal Printer) |
13mm/sec |
|
Display
|
High
definition colour touch screen |
|
Power Supply |
Input:
100-240V AC 50-60Hz Output: 12V 2.5A |
|
Battery Pack
|
Rechargeable
NiMH 8.4V 1Ah |
|
Dimensions
|
25.5cm x
12cm x 3.5cm Transducer 50 x 60 x 90mm |
|
Weight |
Excluding
any transducers : 630g |
|
Temperature
|
The
instrument will operate in a uniform
environment of 0°C - 40°C, out of direct
sunlight. |
|
Operating
Humidity |
30% to 90%
RH |
|
Storage
Temperature |
-20 to +70°
C |
|
Storage
Humidity |
10% to 90%
RH |
|
Connectivity |
RS232 serial
and USB 1.1 |
  
|
|





|